MITECH MHB-3000 liquid-crystal display High accuracy Electronic Brinell Hardness Tester

Categories Digital Hardness Tester
Brand Name: Mitech
Model Number: MHB-3000
Certification: ISO..
Place of Origin: Beijing,China
MOQ: 1 SET
Payment Terms: T/T, Western Union,Paypal
Supply Ability: 5 sets per day
Delivery Time: 8-10 days after paid
Packaging Details: carton box
Measuring range: 8 – 650 HBW
Testing Force Application Mode: Automatic (loading, holding, unloading)
Indenter objective lens conversion mode: manual operation
Display usage: LED eight digital tube display
Microocular magnification: 20X
Test force holding time: 5~60s
Minimum division: 0.005 mm
Size: 753×550×236 mm
Total Weight: 123 kg
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
  • Submit Buying Request
    • Product Details
    • Company Profile

    MITECH MHB-3000 liquid-crystal display High accuracy Electronic Brinell Hardness Tester


    MITECH MHB-3000 Electronic Brinell Hardness Tester

    Overview

    Mitech MHB-3000 Electronic Brinell Hardness Tester, based on the mechanical principle of hard alloy indenter pressing into the sample surface to produce indentation, realizing the material hardness measurement by measuring the diameter of the indentation , it is novel and high reliable with accurately measurement. LED eight digital tubes and luminescent secondary tube display, simple operation, it can visually display the test results to meet the hardness testing requirement for the quality control and qualified assessment of the workpiece sample. It is widely used in metal processing and manufacturing, various metal material’s failure analysis and other fields like colleges and research institutions. It is the new type Brinell hardness testing instrument for testing the hardness of the materials like cast iron, steel, soft alloy and so on.

    Technical Parameters

    Technical specificationsTechnical Parameters
    The power series612.5N(62.5kgf),980N(100kgf),1225N(125kgf),1837.5N(187.5kgf), 2450N(250kgf),4900N(500kgf),7350N(750kgf),9800N(1000kgf),14700N(1500kgf),29400N(3000kgf)
    Measuring range8 – 650 HBW
    Testing Force Application ModeAutomatic (loading, holding, unloading)
    Indenter objective lens conversion modemanual operation
    display usageLED eight digital tube display
    Microocular magnification20X
    Test force holding time5~60s
    Minimum division0.005 mm
    Max sample height220mm
    Max distance from head to body270mm
    VoltageAC 220V/50Hz
    Size753×550×236 mm
    Total Weight123 kg

    Indication accuracy

    Standard BlockIndicating Error%(H)Repeatability Error
    ≤125±30.035`d
    125<HBW≤225±2.50.030`d
    >225±20.025`d

    H:Hardness of standard block `d: Indentation diameter(average)

    Working Principle

    Hardness is not a simple physical quantity, but a reflection of the material elasticity, plasticity, strength and toughness. And hardness test is the simplest mechanical testing method to determine the metal material performance, also one of the important means to judge the products quality.

    Brinell hardness test: Test force (F) is on the steel ball with certain diameter(D) , pressed on sample surface. After a period of time, cancel the force. The indentation diameter is get by measuring with micrometer ocular, thus to calculate the average pressure(N/mm²).Then we can get the Brinell hardness of the sample as below:


    Tips:F—Test force on steel ball,unit: N

    D—Diameter of steel ball ,unit: mm

    d—Indentation diameter,unit: mm

    0.102—Rule coefficient



    Features

    • Suitable for measuring the surface is more rough cast iron, steel and other non-homogeneous specimen Brinell hardness;
    • Using electronic automatic loading system, remove the load weight;
    • With ten test force, a wider range of tests, higher precision;l
    • The control part is the closed-loop control system, which can dynamically reflect real load changes.
    • Modeling novel, strong structure,high reliability and operability , intuitive, high test efficiency;
    • Equipped with excellent performance of the carbide indenter, high hardness, wear resistance, good toughness, while high temperature, corrosion resistance, to ensure that the instrument measured standard, stable and reliable;
    • Host stability is good, the workpiece surface quality and man-made factors on the hardness of the test results less impact;
    • Panel with two-color LED display, red for small gear, green for large-scale force;
    • Compliance with ISO 6506, ASTM E10-12, ASTM E-384, GB / T231.2, JIS Z2243 and other relevant domestic and foreign standards.

    Applied condition

    • Cast iron, steel, non-ferrous metals, soft alloy of metal material (please see table below);
    • Hard plastic, bakelite and some nonmetallic material.
    MaterialsBrinell hardness0.102F/D2
    Steel and cast iron<14010
    ≥14030
    Copper and copper alloys<355
    35~13010
    >13030
    Light metals and their alloys<352.5
    35~805,10
    >8010
    F:Test force(k) D:Head diameter(mm)

    Application field

    • Metal processing industry quality control links
    • Universities teaching and demonstration test
    • The failure test of metal material
    • The material hardness test of scientific research institutions

    Working conditions

    • Working Temperature:18℃~28℃;
    • Relative Humidity:≤65%;
    • Clean environment, no vibration;
    • No corrosive media around.

    Configuration

    Standard

    Configuration


    No.ItemQuantityNote
    1Main body1
    220×Lens1
    3φ2.5mm ball1
    4φ5mm ball1
    5φ10mm ball1
    6Small testing tableDiameter 80mm
    7Large testing table1Diameter 200mm
    8V-shape testing table1For cylindrical sample
    9

    Standard Hardness Block

    HBW/3000/10(150-250)

    1
    10

    Standard Hardness Block

    HBW/1000/10(75~125)

    1
    11

    Standard Hardness Block

    HBW/187.5/2.5(150-250)

    1
    12Fuse wire(2A)3
    13Power Cable1
    14RS232 Cable1
    15Plastic dust cover1
    16Attached files1
    17Instrument case1

    Maintenance and care

    • Each technical requirement of the instrument has already reached the national standards before leaving factory. But its disassembly, transportation or voltage after delivery might cause some data in the instrument changed. So to make the electronic component in normal operation to reduce the testing error, you can use test force of 29400N(3000kgf) pressing times before starting the instrument to eliminate the instrument deformation.
    • The anti-interference component has been installed in the circuit to ensure the transmission of the loading or unloading single of the test force at all levels. At the same time, make sure no strong interference source around when using the instrument, and power off the instrument after finished the testing.
    • All the screws and the caps have been fasten tightly before the instrument leaving factory. However, the screw and cap may become less crowded because that the screw lever for loading and unloading are rotating all the time. If there is irregular noise, when the screw lever rotating, please check and tightened the screw and cap as the below figure shown.

    • The surface of the tested parts should be smooth without dirt, oxide, pits and impressive processing traces;
    • The selection of the test force shall make the indentation diameter within the prescribed range: 0.25D < d < 0.6D (D: indentation diameter, D: steel ball diameter);
    • The minimum thickness of the specimen shall be greater than 10 times of the indentation depth. After testing, the back of the specimen shall not have visible deformation marks;
    • The Instrument power supply shall be grounded reliably and equipped with voltage-stabilizing device;
    • If there is some slight noise during loading or unloading, it is the loading mechanism making adjustment automatically, which is a normal phenomenon.
    • The instrument screw level and other movable surface should be lubricated regular.
    • When not in use for a long time, please cover the dust cover to prevent dust into the instrument interior. And the hardness block, indenter should be coated with antirust oil to prevent rusting after using.
    • Start using again after long time idling, please make testing after preheating the electrical components.
    • The eyepiece must be calibrated to zero for the first time using. No need do zero calibration in subsequent testing.
    • The accuracy of the reading microscope has been adjusted before leaving factory. It is prohibited to disassembly without authorization. Otherwise, it will reduce the reading accuracy of the micro eyepiece.
    • The storage and using of the reading microscope should avoid dust, damp and corrosive gas environment;
    • If there is dirt on the surface of the reading microscope lens, please wipe it with soft absorbent cotton or lens paper. If there is oil pollution, please wipe it slightly by the absorbent cotton with industrial alcohol or ether.
    • Don’t disassemble the instrument without authorization, maintenance related matter, please contact MITECH after-sale service department with 4000600280.

    Quality MITECH MHB-3000 liquid-crystal display High accuracy Electronic Brinell Hardness Tester for sale