Sign In | Join Free | My esadidasol.com
esadidasol.com
Products
Search by Category
Home > Hardness Testers >

Automatic Rockwell Hardness Testing Machine Touch Screen Operation Mitech MHRS-150-XYZ

Categories Digital Hardness Tester
Brand Name: Mitech
Model Number: MHRS-150-XYZ
Certification: EN-ISO,GB,ASTM,JJG...
Place of Origin: Beijing China
MOQ: 1 piece
Payment Terms: T/T, Western Union,PayPal
Supply Ability: 50units per month
Delivery Time: 5-8 work days
Packaging Details: Wooden box
Preliminary testing force: 98.07N,tolerance±2.0%
Indentor specification: Diamond cone Rockwell indenter, Φ1.5875mm steel ball indenter
Display: High sensitivity touch screen
Duration time: 0~99s
Indication error: 0.1HR
Maximum height of specimen: 300mm
Power supply: AC220V/50Hz
Dimension: 550*320*750mm
Main unit weight: 100kg
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
  • Submit Buying Request
    • Product Details
    • Company Profile

    Automatic Rockwell Hardness Testing Machine Touch Screen Operation Mitech MHRS-150-XYZ

    Mitech MHRS-150-XYZ Automatic Rockwell Hardness Tester

    Overview

    Mitech MHRS-150-XYZ automatic Rockwell hardness tester, based on the mechanical principle of conical diamond or hard alloy indenter pressing into the sample surface to produce indentation, realizing the material hardness measurement by measuring the depth of the indentation. With high automation degree, accurate measurement, equipped with automatic turret and high sensitivity touch screen, it can ensure automatic loading and unloading of electronic, automatic indentation marks, automatic generation of inspection reports and other functions. With easy operation and high detection efficiency, it can meet the needs of workpiece sample hardness control and conformity assessment of hardness testing. It is widely used in metal processing and manufacturing, various metal material’s failure analysis and other fields like colleges and research institutions. it is the professional precision testing equipment to improve the work efficiency, product qualification rate and save production costs.

    Technical Parameters

    Technical specificationsTechnical Parameters
    Preliminary testing force98.07N,tolerance±2.0%
    Testing force588.4N,980.7N,1471N,tolerance±1.0%
    Measuring range

    HRA:20-88,HRB:20-100,HRC:20-70,HRD:40-77,

    HRE:70-100,HRF:60-100,HRG:30-94,HRH:80-100,

    HRK:40-100,HRL:50-115,HRM:50-115,HRR:50-115

    Testing force application ModeAutomatic operation (preliminary test force need manual operation)
    Indentor specificationDiamond cone Rockwell indenter, Φ1.5875mm steel ball indenter
    DisplayHigh sensitivity touch screen
    Rockwell scaleHRA,HRB,HRC,HRD,HRE,HRF,HRG,HRH,HRK,HRL,HRM,HRP,HRR,HRS,HRV
    Conversion scaleHRA,HRB,HRC,HRD,HR15N,HR30N,HR45N,HR15T,HR30T,HR45T,HV,HBW,HK
    Duration time0~99s
    Indication error0.1HR
    Maximum height of specimen300mm
    Distance of indenter to outer wall165mm
    Power supplyAC220V/50Hz
    Dimension550*320*750mm
    Main unit weight100kg

    Indication error

    ScaleStandard Hardness RangeAllowed Maximum Tolerance
    HRA20-75HRA±2HRA
    75-88HRA±1.5HRA
    HRB20-45HRB±4HRB
    45-80HRB±3HRB
    80-100HRB±2HRB
    HRC20-70HRC±1.5HRC
    HRD40-70HRD±2HRD
    70-77HRD±1.5HRD
    HRE70-90HRE±2.5HRE
    90-100HRE±2HRE
    HRF60-90HRF±3HRF
    90-100HRF±2HRF
    HRG30-50HRG±6HRG
    50-75HRG±4.5HRG
    75-94HRG±3HRG
    HRH80-100HRH±2HRH
    HRK40-60HRK±4HRK
    60-80HRK±3HRK
    80-100HRK±2HRK
    HRL100-120HRL±1.2HRL
    HRM85-110HRM±1.5HRM
    HRR114-125HRR±1.2HRR

    Features

    • Widely used for high-precision hardness testing for parts with a variety of metal and non-metallic materials;
    • High degree of automation, accurate measurement, suitable for large demand or high precision measurement of high-end users;
    • Automatic operation with one-button, measuring pressure can be adjusted up and down, instant display material hardness value,and high test efficiency;
    • Measurement data files can be set to automatically store, open, store, print, modify, call and other operations, generate word or Excel test report;
    • Adopt touch screen display interface, display operation integration, simple and intuitive, no technical requirements for the operator;
    • Option for various specifications of the indenter, support 15 types of Rockwell hardness scales testing;
    • Equipped with high-speed thermal printer, quickly print out the test data;
    • Support the conversion among various hardness scales such as Brinell, Vickers and etc;
    • Equipped with excellent performance of cemented carbide or diamond indenter, high hardness, wear resistance, good toughness, with high temperature, corrosion resistance, to ensure that the instrument test value is accurate, stable and reliable;
    • With the function of error value correction, the hardness value of the error can be corrected by key input, easier to meet the test accuracy requirements;
    • With the function of threshold overrun automatic alarm, apply to the bulk of finished products or semi-finished pieces of paper-by-piece detection;
    • Original ambient temperature real-time display function, to avoid the instrument working in the case of high or low temperature for a long time, resulting in increased test error and reduce the service life;
    • Consistent with EN-ISO-6508,GB/T230.1,GB/T230.2,JJG112,ASTM E18 and other relevant standards at home and abroad.

    The Scope of Application

    • Sample thickness: the specimen should have a certain size and thickness to ensure the distance between the adjacent indentation center and the distance from the indentation center to edge of the specimen is greater than 3mm, and the minimum thickness of the specimen should not be less than 8 times the depth of indentation. After the test, the back of the sample shall not have obvious deformation marks, and the minimum thickness of the sample depends on the load size used in its material and hardness test. The following table for the minimum thickness of the sample and different hardness test scale can test the material and hardness range table for reference.

    Minimum specimen thickness table

    ScaleRockwell hardness value HRMinimum thickness (mm)ScaleRockwell hardness value HRMinimum thickness (mm)
    A700.7B801.0
    800.5900.8
    900.41000.7
    B252.0C201.5
    301.9301.3
    401.7401.2
    501.5501.0
    601.3600.8
    701.2700.7

    Different hardness test scales can be used to measure the material and hardness range of the sample materials.


    ScaleIndenter typepreliminary testing forceTesting force (N)

    Measuring

    range

    Application
    A

    Diamond

    cone

    98.07(N)588.4(N)20-88HRAhard alloy, carbide, surface quenched steel, carburizing steel
    D980.7(N)40-77HRDthin steel sheet, surface quenched steel
    C1471(N)20-70HRCquenched steel, tempered steel, chilled cast iron
    F

    Φ1.5875

    1/16 inch steel ball


    588.4(N)60-100HRFcast iron, aluminum, magnesium alloy, bearing alloy
    B980.7(N)20-100HRBmild steel, copper alloy, annealed steel
    G1471(N)30-94HRGphosphorus iron, beryllium bronze, malleable cast iron
    H

    Φ3.175

    1/8inch steel ball


    588.4(N)80-100HRHaluminum, zinc, lead etc.
    E980.7(N)70-100HREbearing alloy, tin, hard plastics and other soft materials
    K1471(N)40-100HRK
    L

    Φ6.35

    1/4 inch steel ball

    588.4(N)50-115HRLHard plastic ,hard rubber, aluminum, tin, bronze, mild steel, synthetic resin, friction materials and etc.
    M980.7(N)50-115HRL
    R

    Φ12.7

    1/2 inch steel ball

    588.4(N)50-115HRL

    Note: A, B, C are the commonly used test scales for Rockwell hardness test.

    Applications

    • Used for quality control in metal processing manufacturing;
    • Used for failure analysis testing of metallic materials;
    • Demonstration experiment for education and teaching in Colleges and Universities;
    • Hardness testing of materials in scientific research institutions.

    Working Conditions

    • Operation Temperature:10~30℃;
    • Relative Humidity:≤65%;
    • The surrounding environment should avoid of vibration, strong magnetic field, corrosive medium and heavy dust.

    Maintenance and Precautions

    • The test personnel should follow the operating procedures.Before and after the test,it should be repeated with standard block calibration equipment;
    • If the specimen has visible deformation marks on the back and edges, the test result would be useless. It should choose the intender or test force to test again;
    • Normally, the first point is only used to offset the gap of each support surface without needing to record the test result. Record the statistical test result from the second point;
    • During testing, to ensure accuracy, please make sure the distance between the adjacent indentation center and the distance from the indentation center to edge of the specimen is greater than 3mm;
    • Pay attention to palm the instrument for carrying to forbid falling across. Please fix well the measuring lever, counterweight and boom before carrying;
    • It is strictly prohibited to turn the variable loading handwheel when the instrument load or unload the test force. Rotating the handwheel forcibly will cause the internal gear dislocation, test force in chaos and even damage the internal parts of the instrument;
    • Turn off the power after using the instrument. When not in use for a long time, please cover the dust cover to prevent dust into the instrument interior. And the hardness block, indenter should be coated with antirust oil to prevent rusting after using;
    • Long time after the re-enabled,it should be preheated the electrical components before test
    • The hardness tester should be calibrated at least once a year to ensure its accuracy;
    • Don’t disassemble the instrument without authorization, for maintenance related matters, please contact MITECH after-sale service department with 4000600280

    Quality Automatic Rockwell Hardness Testing Machine Touch Screen Operation Mitech MHRS-150-XYZ for sale
    Send your message to this supplier
     
    *From:
    *To: Mitech CO.,LTD.
    *Subject:
    *Message:
    Characters Remaining: (0/3000)
     
    Inquiry Cart 0